Purge Gases and Flow Rates

The DSC Q Series™ instruments utilize purge gases during experiments.

Select from the following topics for more information:

Choosing the DSC Purge Gases

DSC experiments should always be conducted with a purge gas. With a Mass Flow Controller (MFC), you can switch between two gases for your experiments. Each gas is plumbed separately to the gas ports on the  back of the instrument.

NOTE: Please make sure that you run your experiments with the same gas that you used to calibrate the system. For example, if you calibrate using nitrogen, make your runs with nitrogen.

Back to top

NOTE: Please make sure that you run your experiments with the same gas that you used to calibrate the system. For example, if you calibrate using nitrogen, make your runs with nitrogen.

CAUTION: The mass flow controller is not designed to work with corrosive gases.

WARNING:  Do not use hydrogen or any other explosive gas in the DSC. If you are using samples that may emit harmful gases, vent the gases by placing the DSC near an exhaust.

WARNING:  Do not use any liquid in the purge lines.

Back to top

Setting the Flow Rates

The gas entries on the Notes Page of the Experiment View are used to specify the gas (#1 or #2) and the associated flow rate for each run. These settings will automatically set the gas and flow rate on a run by run basis. The scheduled run's purge settings will take effect immediately once the Apply button is clicked. To change the gas type, access Instrument Preferences.

Recommended Flow Rate:   50 mL/min

The gas flow requirements are different for each method, however, you can use the following guidelines as a starting point to set up the gas purge and air cool flow rates:

*  Use to augment the base purge during conditioning and/or when opening the cell lids.

** The pressure is internally regulated to 15 psi by a valve in the instrument.

Back to top

Switching Gases in Your Experiment

You can set up your method so that the purge gas is switched during the procedure. This is done using the Gas Segment, which controls the switching of the gas between Gas 1 and 2. In addition, you can also change the flow rate within a method using the flow rate method segment.

During an active experiment you can switch gases using the Control/Gas menu or the GAS button on the instrument's Control touch screen.

Back to top

Related Topics