Calibrating the DSC

 

To obtain accurate experimental results you should calibrate the DSC cell when you first install it.  For the best results, you should recalibrate periodically.  The calibration procedures that you need to perform will vary depending on the instrument model and heat flow selection. These experiments will be performed using the instrument control software. See the chart below for a general guide to types of calibration.  For details on how to perform that calibration, refer to linked topics in this section.

The various calibration procedures can be performed several different ways, depending upon the DSC instrument in use.

Calibration consists of several different types of procedures specific to each cell, which are described briefly below.  

Select from the following topics for more information:

NOTE:  At initial installation of the DSC, RCS, or LNCS or after the DSC has been shut down for more than 24 hours. It is recommended that the system be "conditioned" before you start calibration. Click here Setting Up a DSC Cell/Cooler Conditioning Procedure for details on the conditioning procedure.

 

Model

Tzero*

Cell Constant  

Temperature

Baseline

Pressure  

Heat Capacity**

MDSC®

Heat Capacity**

Q1000

DSC Q20/Q10

 

X

X

X

 

 

 

DSC Q200/Q100

Heat Flow T4

X

X

X

 

 

X

 

DSC Q200/Q100

Heat Flow T1

 

 

X

X

X

 

 

 

 

DSC Q1000 (with standard cell)

Heat Flow T4P

X

X

X

 

 

X

X

DSC Q2000/Q1000 (with standard cell)

Heat Flow T4

X

X

X

 

 

X

X

DSC Q2000/Q1000 (with standard cell)

Heat Flow T1

 

X

X

X

 

 

 

 

 

DSC Pressure Cell

 

X

X

X

X

 

 

Back to top

*  Tzero™ describes proprietary technology invented by Robert Danley of TA Instruments—Waters LLC and patented by TA Instruments—Waters LLC (U.S. Patent No. 6,431,747).

** Heat Capacity calibration is NOT performed in cal mode and is not required except when making absolute measurements of heat capacity.

NOTE:  Perform calibration runs that encompass the temperature range you plan to use in your experiments. (We recommend calibration over the widest temperature range.) If your subsequent experiments cover a wider range, you may wish to recalibrate.

NOTE: Please make sure that you run your experiments with the same gas that you used to calibrate the system. For example, if you calibrate using nitrogen, make your runs with nitrogen.

Calibration results may be viewed in the Cell/Temperature Table window and by using the Calibration Report.

Back to top

When to Calibrate

For optimum experimental results you will need to perform all DSC calibrations whenever you change one of the following parameters:

For optimum experimental results you will need to perform all DSC calibrations except Tzero whenever you change one of the following parameters:

Calibration is performed in the instrument’s calibration mode, which is accessed through the Experimental View/Summary Page.  (This does not apply to heat capacity calibration, which is performed in both the standard and MDSC modes, where applicable).

Back to top

Pressure Calibration

Pressure calibration is an optional calibration procedure for the pressure DSC cells.  It is based on comparing the pressure reading at two points, typically one atmosphere and another pressure selected by you, to the pressure reading on an external pressure gauge.

Back to top

Related Topics